When the peak voltage being measured is greater than
three times the RMS AC range selected, UNCAL w ill be
displayed on the LCD readout.
1.
Rotate the RANGE/FUNCTION switch to RMS AC
and select 350 V.
2.
Connect the — (negative) lead to the reference point
(usually a ground or test point) and connect the + (positive)
lead to the unknown voltage to be measured.
3.
Observe the LCD readout. If necessary, select the
next lower value range to obtain maximum resolution
w ithout overranging.
DC Volts
I f the LCD readout exceeds 5 0 0 V, im m ediately
disconnect the test leads fro m the unit-under-test
to prevent possible in stru m e n t damage.
1.
Rotate the RANGE/FUNCTION switch to DC and
select 500 V.
2.
Connect the — (negative) lead to the reference point
(usually a ground or test point) and connect the + (positive)
lead to the unknown voltage to be measured.
3.
Observe the LCD readout. If necessary, select the
next lower value range to obtain maximum resolution
w itho ut overranging.
68
2337 Operators