Using via Digital Data Communication
------------------------------------------------------------------------------------
97
SEQuence:COUNt
:Sets Sequence Operating Number of time
Function:Sets the sequence operating number of time.
Same settings as FUNCTION Settings Item 101
Format:SEQuence:COUNt[]<NRf>
Type:SET command, QUERY command
Parameters:0~9999
Response:Same as parameters
Restrictions: Alarm response is returned for normal SET commands in the following conditions
・While operating as a series operation slave unit
・While operating as a parallel operation slave unit
・
Sequence operation in progress
<Ex.>
SEQ:COUNt []1
SEQ:COUNt?
1
SEQuence:MEMA
:Sets Memory A Operating Time
Function:Sets the memory A operating time
Same settings as FUNCTION Settings Item 102~104
Format:SEQuence:MEMA[]※1,※2,※3
Type:SET command, QUERY command
Parameters:※1 0~999(h)
※2 0~59(m)
※3 0.0~59.9(s)
Response:Same as parameters
Restrictions: Alarm response is returned for normal SET commands in the following conditions
・While operating as a series operation slave unit
・While operating as a parallel operation slave unit
・
Sequence operation in progress
<Ex.>
SEQ:MEMA[]100,10,10.0
SEQ:MEMA?
100,10,10.0
Содержание ZX-S-H Series
Страница 1: ...Constant Voltage Constant Current DC Power Supply series Instruction Manual ...
Страница 37: ...Basic Operations 33 This is blank page ...
Страница 59: ...Useful Function 55 This is blank page ...
Страница 115: ...Using via Digital Data Communication 111 This is blank page ...
Страница 141: ...Controlling with External Analog Contact Signals 137 This is blank page ...
Страница 147: ...Special Loads 143 This is blank page 守 ...
Страница 188: ... 184 付録 External Diagrams ZX S 400H 400HA 4 M3 ...
Страница 189: ...Appendix 185 ZX S 800H 800HA ...
Страница 190: ...Appendix 186 ZX S 1600H 1600HA ...