Controlling with External Analog & Contact Signals
------------------------------------------------------------------------------------
119
Setting Procedures
Press the Function “FUNC.” key.
FUNC. lamp will light.
Settings Item Number (top) and parameters (bottom) will appear in the Number Display.
Select Item number 11 with the VOLTAGE dial.
Change set value with the CURRENT dial.
Parameter=0;
Turns off ON/OFF Toggle at external contact (invalid).
Set if wishing to control by remote command or with the
front panel’s ON/OFF key.
Parameter=1;
Sets Output ON if the external contact is closed.
Sets Output OFF if the external contact is opened.
Output can only be controlled by external contact points.
Power cannot be controlled by remote command (commands)
or by the front panel’s ON/OFF key.
Parameter=2;
Allows the output to be toggled from the panel when
the external contacts are closed.
Also, sets the Output OFF if the external contact is opened.
Output can be controlled by panel or remote command when
external contacts are closed.
Sets output to OFF when contact signal is open
regardless of remote command or panel operations.
Press the FUNC. key again to finalize settings.
Returns the Number Display to measurement display mode.
OUTPUT
PRESET
OVP
OCP
FUNC.
ALARM RESET
CC
CV
OUTPUT POWER (%)
STEP
CURRENT
STEP
VOLTAGE
PRESET FUNC. REMOTE LOCK
20
60
100 LMT
40
80
CC
CV
OUTPUT POWER (%)
STEP
CURRENT
STEP
VOLTAGE
PRESET FUNC. REMOTE LOCK
20
60
100 LMT
40
80
CC
CV
OUTPUT POWER (%)
STEP
CURRENT
STEP
VOLTAGE
PRESET FUNC. REMOTE LOCK
20
60
100 LMT
40
80
CC
CV
OUTPUT POWER (%)
STEP
CURRENT
STEP
VOLTAGE
PRESET FUNC. REMOTE LOCK
20
60
100 LMT
40
80
OUTPUT
PRESET
OVP
OCP
FUNC.
ALARM RESET
Содержание ZX-S-H Series
Страница 1: ...Constant Voltage Constant Current DC Power Supply series Instruction Manual ...
Страница 37: ...Basic Operations 33 This is blank page ...
Страница 59: ...Useful Function 55 This is blank page ...
Страница 115: ...Using via Digital Data Communication 111 This is blank page ...
Страница 141: ...Controlling with External Analog Contact Signals 137 This is blank page ...
Страница 147: ...Special Loads 143 This is blank page 守 ...
Страница 188: ... 184 付録 External Diagrams ZX S 400H 400HA 4 M3 ...
Страница 189: ...Appendix 185 ZX S 800H 800HA ...
Страница 190: ...Appendix 186 ZX S 1600H 1600HA ...