Calibration
------------------------------------------------------------------------------------
160
Offset/Full-Scale Calibration of Output Current with External Analog Values
(ZX-S Series Type-HA)
Output Current Control with External Voltage
1. Perform load with only the shunt resistor.
Disconnect the load wires and the remote sensing wires
.
Connect the shunt to the output terminal. Connect the
digital multimeter to the voltage terminal located on the
shunt.
Turn the POWER switch ON and OUTPUT OFF in
advance.
2. Change the top Number Display to Settings Item 71
with the VOLTAGE dial.
3. Change the set value to 1 (External Voltage) with
the CURRENT dial.
1: External Voltage(0~10V)
4. Press the FUNC. key again to finalize settings.
5. Press the OCP key ①. Set to “44.00A”(※) with the
CURRENT dial ②.
※For 400H/HA set to “5.50A”
For 800H/HA set to “11.00A”
For 1600H/HA set to “22.00A”
6. Input an external voltage of “0.10V”.
IN
OUT
4
20
外部電圧
0~10V
7. Press the OUTPUT key ② to set the output to ON.
CC
CV
OUTPUT POWER (%)
STEP
CURRENT
STEP
VOLTAGE
PRESET FUNC. REMOTE LOCK
20
60
100 LMT
40
80
CC
CV
OUTPUT POWER (%)
STEP
CURRENT
STEP
VOLTAGE
PRESET FUNC. REMOTE LOCK
20
60
100 LMT
40
80
OUTPUT
PRESET
OVP
OCP
FUNC.
ALARM RESET
Note:
Current is 5A for
(400H/HA)
800H/HA is 10A
1600H/HA is 20A
Use wires which will
not produce a voltage
drop below 0.8V.
Calibration Circuit
+
-
Digital Multimeter
Power Supply
V
Shunt
-
+
OUTPUT
PRESET
OVP
OCP
FUNC.
ALARM RESET
External
Voltage
0~10V
Set FUNCTION Settings Item 1, Sink Function, to 1 (1=ON)
and Settings Item 13, Linearity compensation Function, to
0(0=OFF)beforehand.
Caution
OUT PU T
PR ES ET
OVP
OCP
POWE R
FU NC.
LOCAL
LOCK
DISPL AY
A
B
C
ST ORE
MEMOR Y
CC
CV
OUTPU T POWER
(%)
ST EP
CUR RE NT
ST EP
V OL TAGE
PRE SET FUN C. REMOTE LOCK
A LAR M R ESE T
20
60
100 LMT
40
80
①
②
ZX-S-400H
Содержание ZX-S-H Series
Страница 1: ...Constant Voltage Constant Current DC Power Supply series Instruction Manual ...
Страница 37: ...Basic Operations 33 This is blank page ...
Страница 59: ...Useful Function 55 This is blank page ...
Страница 115: ...Using via Digital Data Communication 111 This is blank page ...
Страница 141: ...Controlling with External Analog Contact Signals 137 This is blank page ...
Страница 147: ...Special Loads 143 This is blank page 守 ...
Страница 188: ... 184 付録 External Diagrams ZX S 400H 400HA 4 M3 ...
Страница 189: ...Appendix 185 ZX S 800H 800HA ...
Страница 190: ...Appendix 186 ZX S 1600H 1600HA ...