Calibration
------------------------------------------------------------------------------------
158
Output Voltage Control via External Resistance(C)
1. Disconnect the load and remote sensing wires and
connect the digital multimeter to the output terminal.
Turn the POWER switch ON and OUTPUT OFF in
advance.
2. Change the top Number Display to Settings Item 70
with the VOLTAGE dial.
3.Change the set value to 4 (External Resistance C)
with the CURRENT dial.
4: External Resistance(0~10kΩ C)
4.Press the FUNC. key again to finalize settings.
5. Press the OVP key ①. Set to “704.0V” with the
VOLTAGE dial ②.
6. Set external resistance to 10kΩ.
IN
OUT
19
7
外部抵抗
0~10k
Ω
7. Press the OUTPUT key to set the output to ON.
Calibration Circuit
Digital
Multimeter
CC
CV
OUTPUT POWER (%)
STEP
CURRENT
STEP
VOLTAGE
PRESET FUNC. REMOTE LOCK
20
60
100 LMT
40
80
CC
CV
OUTPUT POWER (%)
STEP
CURRENT
STEP
VOLTAGE
PRESET FUNC. REMOTE LOCK
20
60
100 LMT
40
80
OUTPUT
PRESET
OVP
OCP
FUNC.
ALARM RESET
+
-
V
Power Supply
+
-
OUTPUT
PRESET
OVP
OCP
FUNC.
ALARM RESET
External
Resistance
0~10kΩ
OUT PU T
PR ES ET
OVP
OCP
POWE R
FU NC.
LOCAL
LOCK
D ISPL AY
ST ORE
MEMOR Y
CC
CV
OUTPU T POWER
(%)
ST EP
CUR RE NT
ST EP
V OL TAGE
PRE SET FUN C. REMOTE LOCK
A LAR M R ESE T
20
60
100 LMT
40
80
①
②
ZX-S-4 00 L
A
B
C
Set FUNCTION Settings Item 13, Linearity compensation
Function, to 0 (0=OFF) beforehand.
Caution
Содержание ZX-S-H Series
Страница 1: ...Constant Voltage Constant Current DC Power Supply series Instruction Manual ...
Страница 37: ...Basic Operations 33 This is blank page ...
Страница 59: ...Useful Function 55 This is blank page ...
Страница 115: ...Using via Digital Data Communication 111 This is blank page ...
Страница 141: ...Controlling with External Analog Contact Signals 137 This is blank page ...
Страница 147: ...Special Loads 143 This is blank page 守 ...
Страница 188: ... 184 付録 External Diagrams ZX S 400H 400HA 4 M3 ...
Страница 189: ...Appendix 185 ZX S 800H 800HA ...
Страница 190: ...Appendix 186 ZX S 1600H 1600HA ...