Basic Operations
------------------------------------------------------------------------------------
32
Error Codes
Position
Item
Number
Display
Top
Number
Display
Bottom
Cause
Verify/Recovery
WDT
E001
-
CPU may have performed an
illegal operation
Turn POWER switch OFF and restart
AD_OV
E002
-
Abnormal voltage detected in
the power supply circuit
May be damaged.
Contact Takasago Sales Dept.
OHP
E003
OHP
Over heating protection
is triggered.
●Turn power OFF. Leave device to
cool down. Turn power ON
●Check environmental temperatures
● Verify Fan Motor rotates when
power is turned on. (OUTPUT is
OFF)
OVP
E004
OUP
Over voltage protection was
triggered
Check OVP values.
See page 29 to cancel alarm.
OCP
E005
OCP
Over current protection was
triggered
Check OCP values
See page 31 to cancel alarm
BST
E006
-
Slave unit malfunctioned.
(Displays only on Master
when in Parallel operation)
Check slave unit display
P_ERR
E007
-
May have May have had a
short interruption or voltage
dip at the main power
Turn power switch OFF. Check
device’s power supply.
BST_NRDY
E008
-
Slave unit's input voltage is
abnormal or POWER switch is
OFF
Check whether slave unit’s power
switch is ON
Device Detection
Failure
E009
-
Cannot detect number of
parallel devices
Restart power supply
Max. Device No.
OVER
E010
-
More than 11 parallel devices
are connected
Verify No. of parallel devices.
Restart power supply
Device No.
Disagreement
E011
-
Difference in the current No.
of connected parallel devices
and the previous No. of
parallel devices remembered
by the ZX-S power supply
If the present number of parallel
devices is satisfactory, hold the FUNC.
key for more than 2 seconds and
confirm the detected number of
devices.
TRIP
(1600H/HA)
※Displayed briefly
E012
Shdn
TRIP function was triggered
Cancel by pressing OVP + OCP keys
simultaneously or by the remote control
command ALARM CLEAR.
Communication
Failure
E100
-
Data was not received
Re-verify sent message.
Recovered if normal message is received.
E2P Failure
E110
-
Failure to initialize during startup
Cancel by pressing OVP + OCP keys
simultaneously or by the remote control
command ALARM CLEAR.
Содержание ZX-S-H Series
Страница 1: ...Constant Voltage Constant Current DC Power Supply series Instruction Manual ...
Страница 37: ...Basic Operations 33 This is blank page ...
Страница 59: ...Useful Function 55 This is blank page ...
Страница 115: ...Using via Digital Data Communication 111 This is blank page ...
Страница 141: ...Controlling with External Analog Contact Signals 137 This is blank page ...
Страница 147: ...Special Loads 143 This is blank page 守 ...
Страница 188: ... 184 付録 External Diagrams ZX S 400H 400HA 4 M3 ...
Страница 189: ...Appendix 185 ZX S 800H 800HA ...
Страница 190: ...Appendix 186 ZX S 1600H 1600HA ...