Important functions (selection)
7.3 Calibration and validation functions
SIPROCESS UV600
Operating Instructions, 10/2013, A5E03529375-02
99
The following menu is displayed with the existing calibration gas settings (example).
In order to change the parameters of an existing calibration gas configuration, select them
using the
and
keys and activate them by pressing the
key. In order to
create a new calibration gas configuration, select a vacant location (#4 in the example) and
activate it by pressing the
key.
The following menu is displayed with a calibration gas configuration (example):
Use the
and
keys to select the desired parameter, and activate it by pressing
the
key.
The parameters have the following meanings:
Notation in the menu Function
Name
Name of calibration gas configuration (or number of a vacant location)
Active
Yes means that this calibration gas for calibrations/validations can be
selected
Pump off
Status of sample gas pump during use of the calibration gas
Valve
Digital output which is activated during use of the calibration gas for
controlling a solenoid valve
Use
Settings of calibration gas for measurement