Important functions (selection)
7.3 Calibration and validation functions
SIPROCESS UV600
Operating Instructions, 10/2013, A5E03529375-02
117
The procedure is finished when the message 'Display results?' appears. You can now:
●
Leave the calibration procedure/validation measurement using the
key
●
View the results using the
key.
The following menu is displayed (example):
Function 1 (relative) is not provided for the validation measurement. The list therefore
commences with 2. Meaning of the various parameters:
Display on screen
Description
Date
Current date in representation YY-MM-DD
Time
Current time in 24-hour representation
Setpoint
Setpoint of the used calibration gas
Actual value
Mean value of the measured actual values
7.3.6
Display validation results
This menu can be used to display the data of the last validation measurement.
If the date and time are not displayed, a validation measurement has not yet been carried
out.
Note
Display function
A validation cannot be started from this menu!
Menu call:
Main menu → Calibrations → Calibration → Measured component → Results → Zero/reference
point