Technical data
10.1 Technical data
SIPROCESS UV600
172
Operating Instructions, 10/2013, A5E03529375-02
Materials used in the sample gas path
Table 10- 7
Component
Material
Analyzer unit (sample chamber):
Aluminum or stainless steel mat. no. 1.4404 [3], epoxy
resin
Optical window
CaF
2
or quartz [3], epoxy resin
Gas path, gaskets
FKM (Viton™), PTFE, stainless steel mat. no. 1.4571 [3]
Gas inlet/outlet
PVDF, stainless steel 316 [3]
Moisture sensor
Stainless steel mat. no. 1.4571, platinum, epoxy resin
Diaphragm pump
Pump body
Diaphragm
PVDF
FKM (Viton™), EPDM
[3] Depending on version
Electric inputs and outputs
Table 10- 8
Electric inputs and outputs
Analog output
4, 0 ... 24 mA ; floating (electrically isolated),
residual ripple 0.02 mA, resolution 0.1% (20
μ
A),
accuracy 0.25% of full-scale value, max. load 500
Ω
, max.
output voltage 15 V, adjustable start or fault status
Analog inputs
2, 0 ... 20 mA, reference potential GND,
max. signal strength 30 mA, max. overcurrent
protection ±1 A, input load 50
Ω
, converter accuracy
0.5%, max. voltage ±50 V
Relay outputs
8, with changeover contacts, max. voltage ±50 V
Loading capacity: see table 'Max. load on relay contacts'
Digital inputs
8, switching range 14 ... 42 V (external control voltage),
max. voltage ±50 V
Serial interface
RS485, Ethernet (LAN)
Maximum load on relay contacts
Table 10- 9
Field of application
AC voltage
DC voltage
Current (V
rms
)
Standard
Max. 30 V AC
Max. 48 V DC
Max. 500 mA
CSA
either:
Max. 30 V AC
Max. 48 V DC
Max. 50 mA
or:
Max. 15 V AC
Max. 24 V DC
Max. 200 mA
or:
Max. 12 V AC
Max. 18 V DC
Max. 500 mA