Important functions (selection)
7.5 Parameter assignments
SIPROCESS UV600
138
Operating Instructions, 10/2013, A5E03529375-02
The following menu is displayed (example):
The parameters are selected using the
and
keys and imported with
.
The respective subsequent menu is then displayed with the respective options:
●
Status
Switching on and off of dynamic damping.
–
Status ON means:
the dynamic damping is activated as soon as the difference between two successive
measured values is smaller than the set threshold.
–
Status OFF means that the dynamic damping is not active under any circumstances.
●
Time constant: setting of the desired damping constant
●
Threshold: threshold for activation of dynamic damping with status ON
Figure 7-4
Dynamic time constant (example)
Change the 'Status' parameter by selecting the desired status using the
and
keys and saving using the
key.
The values of the time constants and the threshold can be changed as described in section
Character input (Page 64).