Important functions (selection)
7.5 Parameter assignments
SIPROCESS UV600
Operating Instructions, 10/2013, A5E03529375-02
143
The parameters have the following meanings (the information in parentheses represents the
name of the source):
Current
The analog output permanently outputs the current measured value of
the component
[Name of a]
measuring point
•
If this measuring point is active:
the analog output outputs the current measured value of the
component
•
If other measuring points are active:
the analog output constantly outputs the measured value which was
measured by this measuring point.
1.
Select the component whose measured values are to be output by the analog output. In
order to select an auxiliary value: Select "Auxiliary values" and then the desired value. –
Set "None" in order to switch off the analog output.
2.
Set the desired source using the
,
and
keys.
Setting the 'live' zero
In the first menu, select parameter 2 (Zero). The following menu is displayed (example):
The functional principle 'NAMUR standard' means:
Operating state
Measured values in range 4 to 20 mA
Message of category C or M
Constant 2 mA
Message of category F
Constant 0 mA
Select the desired setting of the zero point using the
,
and
keys.
Note
Possibility of misinterpretation
If the setting '2 mA' or '4 mA' was selected for the zero point, signals below these values may
be interpreted as a fault.