Important functions (selection)
7.3 Calibration and validation functions
SIPROCESS UV600
Operating Instructions, 10/2013, A5E03529375-02
105
Setting the availability of a measured component
This setting can be used to deactivate the use for the associated component without
completely deleting the settings for this component.
Menu call:
Main menu → Calibrations → Settings → Calibration gases → Calibration gas setting → Use →
Measured component→ Active
The following menu is displayed (example):
The parameters have the following meanings:
Yes
This calibration gas is available for calibration procedures and validation
measurements for this component
No
This calibration gas is not available for calibration procedures and validation
measurements for this component
You can then set the availability using the
,
and
keys.
Setting the setpoint for a measured component
This setting specifies the setpoint of the gas for the calibrations for which these calibration
gas settings are used.
Criteria for the setting:
●
The setpoint is usually the actual concentration of the component in the used calibration
gas.
●
However, it is also possible to set a setpoint which deviates from the actual concentration
– e.g. in order to compensate a cross-sensitivity.
It is possible to set different setpoints for different components in the same calibration gas
setting. This is meaningful, for example, if a calibration gas mixture is used which contains
several components.
Menu call:
Main menu → Calibrations → Settings → Calibration gases → Calibration gas setting → Use →
Measured component→ Concentration