Important functions (selection)
7.5 Parameter assignments
SIPROCESS UV600
146
Operating Instructions, 10/2013, A5E03529375-02
The parameters have the following meanings:
System status (xO)
Group message of category x
Module status (Six0)
Status message from module i, message category x
Measured value status
(MVix)
Measured value-based message from module i, message category x
Limit status (MViLIj)
Limit message for measured value i, limit j
Current output (AOiR) Output range status of analog output i
Measuring point status
(MPS)
Status of measuring point switchover
[1]
Measuring point (MPiS) Activation status of measuring point i (control signal for external
valve)
Pump off (BVO1)
Control signal for the gas pump
Sample gas (BVO4)
Control signal for external sample gas valve
Calibration gases
Activation status of a calibration gas (control signal for external valve)
None
Not occupied (digital output is in logical status "0")
[1]
0 = purging time, 1 = measuring time
1.
Select the desired source using the
,
and
keys.
2.
Search for the desired source in the follow-up menus.
Selecting the electronic switching logic
In the previous menu, select the 'Output' parameter. The following menu is displayed
(example):
The parameters have the following meanings:
Direct
The digital output is electronically activated (relay is pulled-up) if the source is
also in the logical activated status (open-circuit principle).
Inverted
The digital output is electronically activated if the source is in the logical
deactivated status (closed-circuit principle).
If the source is logical activated, the switching output is electronically in the
deactivated status (relay dropped-out)