Important functions (selection)
7.3 Calibration and validation functions
SIPROCESS UV600
Operating Instructions, 10/2013, A5E03529375-02
101
The parameters have the following meanings:
Yes
This calibration gas configuration can be used
No
This calibration gas configuration cannot be used. This setting can be used to
completely deactivate a calibration gas configuration without having to delete it.
You can then select the desired status using the
,
and
keys.
Pump mode
The sample gas pump can be switched off automatically if this calibration gas configuration
is active – in other words if the associated calibration gas is flowing into the gas analyzer.
Menu call:
Main menu → Calibrations → Settings → Calibration gases → Calibration gas setting → Pump
off
The following menu is displayed (example):
The parameters have the following meanings:
On
The sample gas pump is switched off automatically if the calibration gas is used
Off
The sample gas pump remains switched on if the calibration gas is used
You can then set the desired status using the
,
and
keys.
Valve control
This function determines which digital output is activated if this calibration gas configuration
is active during a calibration procedure or validation measurement. The flow of calibration
gas can be controlled automatically in this manner.
Menu call:
Main menu → Calibrations → Settings → Calibration gases → Calibration gas setting → Valve