Important functions (selection)
7.3 Calibration and validation functions
SIPROCESS UV600
108
Operating Instructions, 10/2013, A5E03529375-02
The parameters have the following meanings:
Yes
This calibration gas is available for the selected calibration procedures and
validation measurements for the associated component
No
This calibration gas is not available for the selected calibration procedures and
validation measurements for the associated component
Using calibration gas as zero gas
These settings define for which calibrations and validations the associated calibration gas is
available as reference gas.
Menu call:
Main menu → Calibrations → Settings → Calibration gases → Calibration gas setting → Use →
Measured component→ As reference gas
Set the reference gas in the same manner as the zero gas.
7.3.3
Carrying out a manual calibration
Various calibration and validation procedures can be started in the menu branch "Calibration
& validation". The following prerequisites must be observed:
Note
Requirements
•
Only use these procedures if they are correctly programmed.
•
Make sure that the setpoints of the calibration gases are correctly parameterized.
•
Every calibration interrupts measuring mode. Therefore inform affected positions of the
interruption in measuring mode.
•
The calibration is only correct if the sequence is correct:
1. Zero point calibration
2. Reference point calibration