Important functions (selection)
7.3 Calibration and validation functions
SIPROCESS UV600
112
Operating Instructions, 10/2013, A5E03529375-02
The procedure is finished when the message 'Display results?' appears. You can now:
●
Leave the calibration procedure/validation measurement using the
key
●
View the results using the
key.
The following menu is displayed (example):
Meaning of the various parameters:
Display on screen
Description
Relative
Drift between this calibration and the previous calibration
Date
Current date in representation YY-MM-DD
Time
Current time in 24-hour representation
Setpoint
Setpoint of the used calibration gas
Actual value
Mean value of the measured actual values
Absolute
Absolute drift
7.3.4
Display calibration results
You can use this function to display the result of the last calibration of the selected
component or auxiliary value
Menu call:
Main menu → Calibrations → Calibration → Measured component → Results → Zero/reference
point