Important functions (selection)
7.3 Calibration and validation functions
SIPROCESS UV600
110
Operating Instructions, 10/2013, A5E03529375-02
The following calibration functions are possible:
Display in menu
Function
Calib. Zero point
Zero calibration
Calib. Ref. point
Reference point calibration
Calib. Ref. point cell
Reference point calibration with calibration cell (optional, only for analyzers
with calibration cells)
Lin. correction
Linearity correction
Results
Display results of last calibration
Drift reset
Delete saved drift values, i.e. reset to '0'
Note
Calibration sequence
Always calibrate the zero point first and then the full-scale value (reference point), otherwise
the calibration of the full-scale value (reference point) will be inexact!
Use the
and
keys to select the desired calibration or validation function, and
activate this by pressing the
key.
A menu is displayed with the selected calibration or validation function, in this case the
calibration of the full-scale value (reference point).
Carrying out a calibration procedure
Menu call:
Main menu → Calibrations → Calibration → Measured component → Calibration function
The following menu is displayed (example):
Proceed as follows:
1.
Connect the calibration gas to the sample gas inlet of the analyzer.
2.
Select
to start with the calibration.