Startup
5.5 Initial calibration
SIPROCESS UV600
Operating Instructions, 10/2013, A5E03529375-02
51
5.5
Initial calibration
In order to allow measurements, the analyzer must be calibrated. To do this, proceed as
follows:
1.
Check as described in Section 'Displaying/adjusting the setpoints of the calibration gases
(Page 50)' whether the parameterized (displayed) setpoints agree with the concentration
of the calibration gases, and adjust if this is not the case.
2.
Select the menu sequence: M
ain menu → BCU (control unit) → Maintenance →
Calib.&Valid.
A menu is displayed with the existing measured components.
3.
Use the
and
keys to select the desired measured component, and activate
this by pressing the
key.
4.
The following menu is displayed (example):
Note
Calibration sequence
Always calibrate the zero point first and then the full-scale value (reference point),
otherwise the calibration of the full-scale value (reference point) will be inexact!
5.
Use the
and
keys to select the desired calibration or validation function, and
activate this by pressing the
key.