Important functions (selection)
7.5 Parameter assignments
SIPROCESS UV600
Operating Instructions, 10/2013, A5E03529375-02
151
7.5.9.2
Configuring a Modbus connection
Menu call:
Main menu → Parameters → Communication → Modbus
The following menu is displayed (example):
You can then change the Modbus parameters as follows:
●
Use the
,
and
keys to select the associated parameter
●
Change the parameters in the follow-up menu according to your requirements.
You can configure the following Modbus parameters:
Notation in the menu
Function
On/off
Modbus active/inactive
Slave address
Device address in accordance with Modbus specification
Type
Interface type (TCP or RTU)
TCP parameters
TCP port
RTU parameters
Baud rate, parity, data bits, stop bits
7.5.10
Setting the date and time
The current date is set in this menu.
Note
Date and time setting
The date and time are retained for approximately 3 to 5 days following disconnection of the
device from the power supply.
Menu call:
Main menu → Parameters → Date
-
time → Date