Important functions (selection)
7.2 Diagnostics functions
SIPROCESS UV600
Operating Instructions, 10/2013, A5E03529375-02
79
The menu contents show the possible auxiliary values, in this case the flow rate with the
current measured value.
7.2.6
Results of calibrations and validations
Display calibration results
This function can be used to display the last calibration results for each individual component
and the associated drift values.
Menu call:
Main menu → Diagnostics → Calibration results.
A menu with the data of the last calibration is then displayed.
1.
Use the
and
keys to navigate to the desired component or auxiliary value,
and select using the
key.
2.
Then select one of the parameters 'Zero' 'Ref. point'.
The following menu is now displayed (example):
The following parameters can be displayed:
Display
Description
Relative
Drift between this calibration and the previous calibration
Date
Current date (representation: DD-MM-YY)
Time
Current time (24-hour representation)
Setpoint
Setpoint of the used calibration gas
Actual value
Actual value of the used calibration gas (mean value)
Absolute
Absolute drift, i.e. total change in a drift value over several calibrations