Mounting and connecting
4.5 Electrical connections
SIPROCESS UV600
36
Operating Instructions, 10/2013, A5E03529375-02
Plug connectors X4 and X5 – digital outputs (DO)
Plug connectors X4 and X5 – digital outputs (DO):
Figure 4-7
Plug connector X4/X5 - digital outputs, representation in de-energized state
Digital outputs DO1 ... DO4 are located on plug connector X4;
digital outputs DO5 .... DO8 are located on plug connector X5
Characteristic of digital outputs (see also Auto-Hotspot):
●
Floating relay changeover contacts
●
Single-pole changeover switch, three connections
●
Maximum voltage: ±50 V
●
Only connect inductive loads (e.g. relays, solenoid valves ...) via spark-quenching diodes
●
Observe maximum loading capacity (see following table)
Table 4- 1
Maximum load per relay changeover contact referred to GND/enclosure
Field of application
AC voltage (rms value)
DC voltage
Current
Standard
Max. 30 V AC
Max. 48 V DC
Max. 500 mA
CSA (possible voltage/current combinations
in CSA standard range)
Max. 30 V AC
Max. 48 V DC
Max. 50 mA
Max. 15 V AC
Max. 24 V DC
Max. 200 mA
Max. 12 V AC
Max. 18 V DC
Max. 500 mA