Important functions (selection)
7.2 Diagnostics functions
SIPROCESS UV600
Operating Instructions, 10/2013, A5E03529375-02
71
7.2.3
Status
Overview of the status messages
Every module and every measured value can assume further statuses in addition to the
'MEASURE' status. Status flags are used to identify the corresponding status:
Status
(flag)
Description
Possible cause
F
Fault
Internal fault/malfunction (unsafe operating state)
M
Maintenance
Maintenance request/maintenance status
[1]
C
Check
An internal function is currently running, e.g. a calibration. Measuring function is interrupted
[1]
U
Uncertain
Unsafe operating state or measured value
[1]
L
Limit
Limit
[2]
T
Timeout
Internal timeout of measured signals, e.g. during a calibration procedure
[2]
E
Extended
Internal message
[2]
Status flags
[1]
Corresponds to specification in accordance with NAMUR
[2]
Without NAMUR significance
The status flags are displayed in the "Status" menu.
The status flags are displayed in the sequence F - M - C - U - L - T - E. A non-active flag is
displayed by an underscore (_), i.e. the display "_ _ _ _" means that no flags are active.
Check status
Menu call:
Main menu → Diagnostics → Status.
The following menu of the function groups with their group status is then displayed
(example):
Select the desired function group in order to check the individual status.