Important functions (selection)
7.3 Calibration and validation functions
SIPROCESS UV600
Operating Instructions, 10/2013, A5E03529375-02
113
The following menu is displayed (example):
Meaning of the various parameters:
Display on screen
Description
Relative
Drift between this calibration and the previous calibration
Date
Current date in representation YY-MM-DD
Time
Current time in 24-hour representation
Setpoint
Setpoint of the used calibration gas
Actual value
Mean value of the measured actual values
Absolute
Absolute drift
7.3.5
Carrying out a manual validation measurement
A validation measurement is a measuring procedure with a calibration gas. A validation
measurement is executed just like a calibration procedure; the results are evaluated as with
a calibration procedure (including drift calculation) and saved.
The calibration is not changed by this.
Validation measurements can only be carried out for measured components, not for auxiliary
values or virtual components.
Starting the validation procedure
Menu call:
Main menu → Calibrations → Validation
A menu is displayed with the current measured components.
Use the
and
keys to select the desired measured component, and activate this
by pressing the
key.