Technical data
10.1 Technical data
SIPROCESS UV600
170
Operating Instructions, 10/2013, A5E03529375-02
Electrical characteristics
Table 10- 3
Electrical characteristics
Power supply
(optional, see rating plate)
93 ... 132 V AC
186 ... 264 V AC
Line frequency (AC)
47 ... 63 Hz
Permissible overvoltages (transients in the power
supply network)
Up to overvoltage category II in accordance with
IEC 60364-4-443
Power consumption (analyzer)
Approx. 50 VA, max. 300 VA
Electrical safety
Tested in accordance with EN 61010-1
EMC– interference immunity
(electromagnetic compatibility)
In accordance with EN 61326-1, EN 61326-2-1,
EN 61000-6-2, EN 61000-6-4 and EU directive
2004/108/EU.
With irradiation of electromagnetic fields in the
frequency range 750 MHz ± 20 MHz, increased
errors in measurement may occur with small
measuring ranges
Internal line fuse
- Primary
- Secondary
6.3 A, not replaceable
8 A
Measuring behavior
Table 10- 4
Measuring behavior
(relating to sample gas pressure 1013 hPa absolute, 0.5 l/min sample gas flow and 25 °C ambient
temperature)
Reference point drift
< ±1%/week of respective span
Zero drift
- Standard measuring ranges
- Small measuring ranges
(≤
2x smallest measuring range)
- Measured components NO, NO
2
, SO
2
< ±1%/week of respective span
< ±2%/week of respective span
< ±1%/day of respective span
Repeatability
< ±1% of respective span
Linearity deviation
< ±1% of respective span
Detection limit (2
σ
)
- Standard measuring ranges
- Small measuring ranges
(≤
2x smallest measuring range)
< 0.5% of span
< 1% of span
Influence of ambient temperature:
- Standard measuring ranges
- Small measuring ranges
(≤
2x smallest measuring range)
< 1%/10 K
< 2%/10 K
Setting time (T
90
)
4 s
Starting time
< 60 minutes