Startup
5.3 Startup
SIPROCESS UV600
44
Operating Instructions, 10/2013, A5E03529375-02
5.2.3
Last measures prior to startup
Check all connections
Following a successful leak test, check all gas inlet and outlet lines and electrical
connections for correct seating.
Read the operating instructions
Before switching on the analyzer, make yourself acquainted with the basic operation of the
analyzer (see Section Operation (Page 53))
.
5.3
Startup
Switch on the power supply. The "POWER" LED will light up.
Wait until the measured-value display appears.
Up to the end of the warming-up phase, the LED "MAINTEN. REQUEST" is also lit.
The startup phase takes between 45 minutes and 2 hours depending on the ambient
temperature. The full measuring accuracy is not achieved during this time, and calibration is
not meaningful.
The display illumination may switch off automatically during this time. To reactivate it, press
the right and left blue function keys simultaneously.
If the LED "MAINTEN. REQUEST" goes out after the startup phase, the basic setting of the
device can be carried out. To do this you must be acquainted with operation as described in
section Operation (Page 53).