Important functions (selection)
7.5 Parameter assignments
SIPROCESS UV600
Operating Instructions, 10/2013, A5E03529375-02
135
Range selection
The following output ranges are possible for the "Range selection" parameter:
0
Output range 0 permanently activated
1
Output range 1 permanently activated
auto
Autoranging
Autoranging responds as follows:
●
If the current measured value is in the small output range (in the output range with the
smaller span), the small output range is automatically active.
●
As soon as the measured value exceeds the full-scale value of the small output range,
the large output range is activated automatically (i.e. the electronic signal span of the
analog output is adapted).
●
As soon as the actual measured value in the large output range falls below 90% of the
full-scale value of the small output range, the small output range is activated again.
Digital output for status of output range
By selecting the "Status" parameter you access the following menu (example):
Use the
and
keys to select the desired digital output. If you subsequently wish to
change the switching logic, press the
key to branch to the follow-up menu.
Set the desired switching logic (direct or inverted) there using the
,
and
keys.