Important functions (selection)
7.2 Diagnostics functions
SIPROCESS UV600
Operating Instructions, 10/2013, A5E03529375-02
89
7.2.12
Service information
This function provides information on the internal measured signals and control values.
These displays may help the the vendor's servicing department when analyzing function
faults. This function is only available in the user level 'Authorized user'.
Menu call:
Main menu → Diagnostics → Service
The following menu is then displayed (example):
Use the
,
and
keys to select a module and to obtain its internal parameters.
The following parameters are possible:
Notation in the menu
Explanation
Electronics temperature * Current temperature of the module electronics
Name/unit
Name of measured component/physical unit of measured value
Phys. range
Physical measuring range of component
Calibration cell *
Basic values of calibration cell
Drift limit
Programmed drift limits
Raw values
Internal values of measured signal processing
A/D values
Internal values of measured signal processing
Heaters *
Setpoints and actual values of internal heater
Motors *
Operating state of an internal motor
Restart
Restart of module electronics
Maintenance flag
Activation of message "M" (status display) for the module
* Optional parameters