HydroRanger 200 – INSTRUCTION MANUAL
Page 165
mmm
mm
Pa
ra
m
e
te
rs
Open Channel Monitoring (P600 to P621)
[6 relay model]
Open Channel Monitoring (OCM) operation is a feature of the HydroRanger 200 6 relay
model only.
If the HydroRanger 200 is used to monitor open channel flow, alter the following
parameters as required and run a calibration as described in P621.
The HydroRanger 200 measures head as referenced to Empty (P006) or OCM Zero Offset
(P605), when Operation is set for OCM (P001 = 6). Flowrate, based on head (at the point of
measure specified by the Primary Measuring Device fabricator) is also calculated and
displayed on the LCD.
Some Primary Measuring Devices require a longer Range Extension (P801) to avoid
entering the LOE failure state if the water level falls below the zero point of the Primary
Measuring Device. See
P801 Range Extension
on page 200 for more information.
P600 Primary Measuring Device [6 relay model]
The type of primary measuring device (PMD) used.
Note:
See
Open Channel Monitoring
(OCM) on page 71 for application examples
involving common weirs and flumes.
Primary Index
Single Point Model
Dual
Global
Transducer
Values
0
* off (no calculation)
1
Exponential (see P601)
2
Palmer-Bowlus Flume (see P602)
3
H-Flume (see P602)
4
Universal Linear Flow Calculation (see P610, P611)
5
Universal Curved Flow Calculation (see P610, P611)
6
BS-3680/ISO 4359 Rectangular Flume (see P602)
7
BS-3680/ISO 1438/1 Thin Plate V-Notch Weir (see P602)
Alters
•
P601 Flow Exponent
•
P602 Primary Measuring Device Dimensions
•
P608 Flowrate Units
Altered By
•
P001 Operation