Page 122
HydroRanger 200 – INSTRUCTION MANUAL
7ML19985FC03
mmm
mm
Pa
ra
m
e
te
rs
Use these parameters to enable the HydroRanger 200 to show readings based on vessel
or wet well volume (rather than level).
P050 Tank Shape [6 relay model]
Enter the Tank Shape value matching the monitored vessel or wet well.
When Operation is
LEVEL
(P001 = 1), liquid (material) volume is calculated.
Alternatively, when Operation is
SPACE
(P001 = 2), remaining vessel capacity is
calculated. In RUN mode, readings are displayed in percent of maximum volume. To
convert readings to volumetric units, see
Maximum Volume (P051)
.
Primary Index
Level