7ML19985FC03
HydroRanger 200 – INSTRUCTION MANUAL
Page 191
mmm
mm
Pa
ra
m
e
te
rs
P728 Shot Delay
The delay, in seconds, between transducer shots.
Use this if transient acoustic noise within the vessel is causing measurement difficulties
due to echoes from one shot being received on the next. If more than one ultrasonic unit
is installed for redundancy, this value should be 0.
P729 Scan Time
View the elapsed time (in seconds) since the point displayed was last scanned.
This may be viewed as an Auxiliary Reading in the RUN mode.
Display (P730 to P739)
P730 Auxiliary Reading
Use this feature to display operator selected Auxiliary Readings temporarily or
indefinitely (as desired).
Select OFF to display Auxiliary Readings temporarily. Select HOLd to display Auxiliary
Readings until another Auxiliary Reading is selected or programming mode is entered.
See the
Hand Programmer
section on page 26 for RUN mode auxiliary readings.
Selecting the Auxiliary Reading operation
1.
Press READING
to display the Auxiliary Function symbol.
2.
Press ARROWS
to access the OFF or HOLd option desired.
3.
Press ENTER
Primary Index
Transducer
Values
Range: 0.1 to 4.0
Preset: 0.5
Primary Index
Level
Values
Range: 0.000 to 9999 (view only)
Related
•
P001 Operation
Primary Index
Global
Values
Range: 000 to 999
Display: OFF, HOLd
Содержание HydroRanger 200
Страница 1: ...Instruction Manual July 2009 200 hydroranger ...
Страница 232: ...Page 222 HydroRanger 200 INSTRUCTION MANUAL 7ML19985FC03 mmmmm Parameters Notes ...
Страница 264: ...Page 254 HydroRanger 200 INSTRUCTION MANUAL 7ML19985FC03 mmmmm Appendix H Software Rev Notes ...
Страница 278: ...Notes ...
Страница 279: ......