ADVIA 2400 Operator’s Guide
Reviewing the calibration
NOTE:
For detailed information on how to use the View Calibration Curve and ISE
Monitor windows, refer to the online Operator’s Guide.
•
Use the View Calibration Curve window to review the calibration of photometric
methods.
•
Use the ISE Monitor window to review the calibration of electrolyte methods.
Calibration/RBL History Window
Use the Calibration/RBL History window to review the calibration history for all
methods that are run at the system and to create CSV files.
1.
At the Menu Panel, select Calibration, then select Calibration/RBL History.
2.
Select the calibration data to display by selecting Latest or Select:
•
Latest
–Displays the most recent calibration data for each method on the system.
•
Select
–Activates the Test Name box to allow a more detailed test selection.
3.
If the Select option is chosen, then select one of the 3 Test Name options:
•
Photometric
–Displays all calibration history data for all photometric methods.
•
ISE Tests
–Displays all calibration history data for all ISE methods.
•
Other
–Activates the test list box to allow specific test selection (highlight each
test to display).
4.
Specify a date range, to display data for calibrations performed on specific days for
the tests selected:
a.
Type the Start Date (yyyymmdd) of the range (if other than the default, which is
the current date).
b.
Type the End Date (yyyymmdd) of the range (if other than the default, which is
the current date).
c.
Select S
earch
to display the calibration data for the tests selected for the
specified date range.
NOTE:
Select and drag the column heading borders to resize the columns in the
calibration data table.
d.
To clear the data for the method selected, select
Clear
.
e.
When prompted, select
Yes
.
5.
To reset the calibration select the
Reset
check box.
NOTE:
You cannot cancel the reset once you select
Save
. If you resize the reset
column an save your changes, those changes are retained when you exit out and back
into the window.
6.
Select
P
rint, to print the data displayed.
7.
To create a CSV file of the displayed data for archiving, or for exporting the data to
another application, do the following:
a.
First, select the data you want to save.
Содержание ADVIA 2400
Страница 1: ...ADVIA 2400 Chemistry System Operator s Guide 073D0403 01 Rev A 2009 06 ...
Страница 12: ...12 ADVIA 2400 Operator s Guide ...
Страница 64: ...64 ADVIA 2400 Operator s Guide ...
Страница 98: ...98 ADVIA 2400 Operator s Guide ...
Страница 165: ...Maintenance 165 ...
Страница 167: ...Troubleshooting 167 ...
Страница 173: ...System Setup 173 ...
Страница 177: ...System Setup 177 ...
Страница 181: ...Appendix 181 ...
Страница 182: ...182 ADVIA 2400 Operator s Guide ...