Maintenance
129
1.
Lift the cover from the saline diluent bottle (
3
), and remove the bottle.
CAUTION
Note the bottle position on the shelf, to avoid mixing up the fluid bottles.
2.
Empty the remaining contents of the bottle.
3.
Rinse the bottle with deionized water and drain well.
4.
Refill the bottle with 0.9% saline diluent.
5.
Replace the bottle in the same position on the shelf in the cabinet.
6.
Replace the cover of the diluent bottle.
NOTE:
Make sure that the Teflon tube and filter holder are located at the bottom of
the dilution bottle.
7.
Prime the fluid lines:
NOTE:
If you are cleaning the detergent or cell conditioner bottles at this time, you
can prime all the fluid lines at once.
a.
At the Operation Panel, select Prime.
b.
At the PRIME Settings dialog box, select Prime 2, then
type
10 or more in each
of he number of times fields.
c.
Select Execute.
Cleaning and replenishing the cuvette wash bottle
Materials required:
• Deionized water
• Cuvette detergent (wash solution)
Time: 10 minutes
Analyzer mode: READY
BIOHAZARD
Wear personal protective
equipment.
Use universal precautions.
1)
ISE buffer bottle
2)
RRV (Reaction) bath oil
3)
Isotonic saline diluent bottle
4)
Cuvette detergent bottle
5)
Cell conditioner bottle
Figure 5-18. Cuvette wash bottle
Содержание ADVIA 2400
Страница 1: ...ADVIA 2400 Chemistry System Operator s Guide 073D0403 01 Rev A 2009 06 ...
Страница 12: ...12 ADVIA 2400 Operator s Guide ...
Страница 64: ...64 ADVIA 2400 Operator s Guide ...
Страница 98: ...98 ADVIA 2400 Operator s Guide ...
Страница 165: ...Maintenance 165 ...
Страница 167: ...Troubleshooting 167 ...
Страница 173: ...System Setup 173 ...
Страница 177: ...System Setup 177 ...
Страница 181: ...Appendix 181 ...
Страница 182: ...182 ADVIA 2400 Operator s Guide ...