122
ADVIA 2400 Operator’s Guide
3.
Replace the reservoir cover.
4.
Replace the lamp access cover.
NOTE:
If adding coolant does not clear the lamp coolant warning, call your local
technical support provider or distributor.
Checking lamp energy
NOTES
Check the lamp energy after cleaning or replacing cuvettes, and after replacing the lamp.
When completing weekly maintenance, be sure to perform the procedures in the
following sequence:
1.
Weekly wash
2.
Lamp energy check (this procedure)
3.
Cuvette blank measurement
Materials required:
No materials required
Time: 15 minutes
Analyzer mode: READY
BIOHAZARD
Wear personal protective equipment.
Use universal precautions.
IMPORTANT
Do not touch or turn the reaction tray at any time during the lamp energy check
procedure. The reaction tray should turn freely. If the reaction tray is shifted, repeat the
procedure, since a shift could result in an erroneous lamp energy reading.
1.
At the Menu Panel, select
Maint
, then select
Lamp Energy Monitor
.
The Lamp Energy Monitor window displays.
2.
Ensure the bottle at position #50 in Reagent Tray 1 (RTT1) contains deionized water.
3.
Select
Check Energy
.
The Lamp Energy Monitor dialog box displays.
4.
Type
50
in the RTT1 bottle posi. field, then select
3: 70 mL
for the Container
field.
5.
Select
Meas. Start
.
•
The reagent probe aspirates deionized water from RTT1 and dispenses it into
reaction cuvette #1.
•
The reaction disk rotates until cuvette #1 is in the detection position.
•
The Operation window displays Lumi.Check and then WAIT.
NOTE:
Perform steps 6–10 while in the WAIT state.
Содержание ADVIA 2400
Страница 1: ...ADVIA 2400 Chemistry System Operator s Guide 073D0403 01 Rev A 2009 06 ...
Страница 12: ...12 ADVIA 2400 Operator s Guide ...
Страница 64: ...64 ADVIA 2400 Operator s Guide ...
Страница 98: ...98 ADVIA 2400 Operator s Guide ...
Страница 165: ...Maintenance 165 ...
Страница 167: ...Troubleshooting 167 ...
Страница 173: ...System Setup 173 ...
Страница 177: ...System Setup 177 ...
Страница 181: ...Appendix 181 ...
Страница 182: ...182 ADVIA 2400 Operator s Guide ...