Operating the System
71
(1) At the Menu Panel, select
Reagent
, then select
Reagent Inventory
.
(2) Execute a Barcode Scan at the Reagent Inventory window.
(3) Evaluate calibration status.
4. Perform a start-up wash.
WASH 3 or Startup wash, 26 minutes
CTT position 51
DI water (tube)
RTT1 position 50
DI water (70-mL container)
RTT2 position 50
DI water (70-mL container)
WASH 2 or Shutdown wash, 38 minutes
CTT position 15
ISE detergent (cup)
CTT position 16
DI water (tube)
CTT position 49
10% cuvette wash (tube)
CTT position 50
DI water (tube)
RTT1 position 49
10% cuvette wash (70-mL
container)
RTT1 position 50
DI water (70-mL container)
RTT2 position 49
10% cuvette wash (70-mL
container)
RTT2 position 50
DI water (70-mL container)
Weekly wash set up
Same as WASH 2, except replace the 10% cuvette wash with 5%
reagent probe wash in both the CTT and RTT positions.
Times required to perform prime, washes, and cell blank
Typical times
Operation Cycles
Time
(minutes)
PRIME1 5 1:40
WASH1 1 9:10
WASH2 2 38
WASH3 1 26
Cell blank
NA
14
Times for combined operations during startup and shutdown
Startup
•
WASH3 = 28 minutes
•
WASH3 + Cell blank = 42 minutes
Содержание ADVIA 2400
Страница 1: ...ADVIA 2400 Chemistry System Operator s Guide 073D0403 01 Rev A 2009 06 ...
Страница 12: ...12 ADVIA 2400 Operator s Guide ...
Страница 64: ...64 ADVIA 2400 Operator s Guide ...
Страница 98: ...98 ADVIA 2400 Operator s Guide ...
Страница 165: ...Maintenance 165 ...
Страница 167: ...Troubleshooting 167 ...
Страница 173: ...System Setup 173 ...
Страница 177: ...System Setup 177 ...
Страница 181: ...Appendix 181 ...
Страница 182: ...182 ADVIA 2400 Operator s Guide ...