74
ADVIA 2400 Operator’s Guide
•
In the Test table, select each test or ratio you want to run.
•
In the Test-tbl no. box, enter the number of the test you want followed by a
period (.).
•
In the Profiles area, select each profile you want to run.
3.
Select
Batch Entry
.
4.
Select
Samp.no.
,
Posi.no
., or
Batch entry
button, then enter corresponding
information in the selected box.
5.
Select
Execute
.
NOTE:
The Posi.no. and Samp.no. fields increment by the number of workorders
requested from Batch Entry.
Deleting or changing test selectivity for multiple workorders
1.
At the Menu Panel, select
Request
then select
Order Entry
.
2.
Select
Batch Func
.
3.
Perform any of the following tasks at the window:
a.
To erase workorders, select
Delete workorder
.
b.
To change the test selectivity, select
Test correct
.
4.
In the Entry format area, select the means of identifying the first workorder in step 6.
5.
In the Last no. entry format
area, select the means of determining the last workorder
in step 7.
6.
In the Start no. box, identify the first workorder you want changed.
7.
In the Last no. box, identify the last workorder you want changed or type the number
of workorders you want changed.
8.
If Testcorrect was selected in step 3, select
Test Table
.
a.
To add a test, select it.
The check mark must appear bold.
b.
To delete a test, double-select it. The check mark must appear dim.
9.
Select
Return
.
10.
Select
Execute
.
Содержание ADVIA 2400
Страница 1: ...ADVIA 2400 Chemistry System Operator s Guide 073D0403 01 Rev A 2009 06 ...
Страница 12: ...12 ADVIA 2400 Operator s Guide ...
Страница 64: ...64 ADVIA 2400 Operator s Guide ...
Страница 98: ...98 ADVIA 2400 Operator s Guide ...
Страница 165: ...Maintenance 165 ...
Страница 167: ...Troubleshooting 167 ...
Страница 173: ...System Setup 173 ...
Страница 177: ...System Setup 177 ...
Страница 181: ...Appendix 181 ...
Страница 182: ...182 ADVIA 2400 Operator s Guide ...