Maintenance
123
6.
At the
Lamp Energy Monitor window, in the Luminous Energy Check area, enter the
settings:
a.
Type
1000
in the Meas. times field.
Enter the number of times to measure the lamp energy (normal setting: 1000).
b.
Type
100
in the Meas. cycle field.
Enter the time (in µs) to elapse after each lamp energy measurement (normal
setting: 100).
c.
Select
AD
.
d.
Select
Auto
.
7.
Select
Meas. Energy
.
The message, "Execute the lamp energy check?" displays.
8.
Select
OK
.
9.
On the Lamp Energy Monitor window, select
Collect Data
.
10.
Calculate the scatter plot:
a.
Note the value of the 340-nm AD count field.
b.
Add 50 to the 340-nm AD count and type the sum in the top field to the left of
the graph, then select
Enter
.
c.
Subtract 50 from the 340-nm AD count (
noted in step 10 a
) and type the
difference in the bottom field, to the left of the graph, then press
Enter
.
The lamp energy displays as a scatter plot.
•
If the points are mostly within
±
40 of the center, the lamp is normal.
•
Otherwise replace the lamp
11.
Only
if you replaced the lamp or a cuvette segment, select
Regist Data
, then select
OK
in the Registration window.
If not, proceed to step 12.
NOTE:
The system uses the data from the lamp energy data registration as the
comparison standard for the next calculation of the attenuation ratio.
12.
Exit the Lamp Energy Monitor window, then select
Initialize
to switch the system
from the WAIT
state to the READY
state.
13.
Run cell blank measurement.
Reading lamp energy data
1.
Execute the lamp energy check and acquire the data.
•
A graph displays at the window.
•
The lamp energy check data displays to the left of the graph.
2.
Select a wavelength for which to display data.
3.
Display the voltage or the AD value.
4.
(Optional) Change the vertical scale of the graph.
Содержание ADVIA 2400
Страница 1: ...ADVIA 2400 Chemistry System Operator s Guide 073D0403 01 Rev A 2009 06 ...
Страница 12: ...12 ADVIA 2400 Operator s Guide ...
Страница 64: ...64 ADVIA 2400 Operator s Guide ...
Страница 98: ...98 ADVIA 2400 Operator s Guide ...
Страница 165: ...Maintenance 165 ...
Страница 167: ...Troubleshooting 167 ...
Страница 173: ...System Setup 173 ...
Страница 177: ...System Setup 177 ...
Страница 181: ...Appendix 181 ...
Страница 182: ...182 ADVIA 2400 Operator s Guide ...