Maintenance
133
4.
Remove the bottle.
CAUTION
Make a note of the bottle position on the shelf, to avoid mixing up the fluid bottles.
5.
Empty any remaining contents of the bottle.
6.
Rinse the bottle with deionized water and drain well.
CAUTION
Ensure that the level-sensor connector does not get wet to avoid damaging it.
7.
Refill the bottle with cuvette conditioner.
8.
Return the bottle to the same position on the shelf in the cabinet.
9.
Connect the cuvette-conditioner bottle level sensor connector, then push the
connector in and turn clockwise.
10.
Insert the filter and hose, then fasten the cap.
NOTE:
Make sure that the filter holder is located at the bottom of the bottle.
11.
Prime the fluid lines:
NOTE:
If you are cleaning other bottles, wait to perform this step for all fluid lines.
a.
At the Operation Panel, select Prime.
b.
At the Prime Settings dialog box, select Prime 2 and type 10 for the Number of
times in all fields.
c.
Select Execute.
Every 3 months maintenance
Replacing the lamp
Materials required:
• Halogen lamp, 12 V/50 W
(REF 02127928, PN 073-0099-01)
Time: 60 minutes
Analyzer mode: OFF
BIOHAZARD
Wear personal protective
equipment.
Use universal precautions.
Содержание ADVIA 2400
Страница 1: ...ADVIA 2400 Chemistry System Operator s Guide 073D0403 01 Rev A 2009 06 ...
Страница 12: ...12 ADVIA 2400 Operator s Guide ...
Страница 64: ...64 ADVIA 2400 Operator s Guide ...
Страница 98: ...98 ADVIA 2400 Operator s Guide ...
Страница 165: ...Maintenance 165 ...
Страница 167: ...Troubleshooting 167 ...
Страница 173: ...System Setup 173 ...
Страница 177: ...System Setup 177 ...
Страница 181: ...Appendix 181 ...
Страница 182: ...182 ADVIA 2400 Operator s Guide ...