ADVIA 2400 Operator’s Guide
Monitoring an analysis while it is running
1.
At the Menu Panel, select
Request
, then select
Test Result Monitor
.
2.
Monitor any of the following at this window:
a.
processing status
b.
sample information and system status
c.
result data information
When the run is complete, the operating mode display becomes END, then returns to
READY if you ran the samples from the STT. The system returns to WAIT if you ran
samples from the rack handler or lab automation system.
3.
Close the window.
Processing status
The center of the window resembles the sample tray. The outer ring is the STT tray and
the inner ring is the CTT tray. Each tray position is a circle.
As the run continues, the circles containing samples change color. The colors indicate the
current status of each sample. The color codes are displayed at the lower-left of the
window.
For additional information about the color codes see page 41.
Sample information and system status
The button bar at the top of the window displays the Sample Search and Rack or LAS.
Sample Info. buttons.
•
Select
Sample Search
to display a dialog box where you can search for sample
information by sample number by STT position (tray number and sample position),
or by rack/LAS position (rack number and sample position).
The search returns a Sample Information window showing the sample number,
position number, sample status, and the time remaining to complete the analysis of
the sample.
•
If a rack handler or laboratory automation system (LAS) is in use, select the
Rack
or
LAS. Sample Info.
button to display the following sample information in the Rack or
LAS. Sample Information window:
♦
Sample barcode number.
♦
Sample status (see the status codes listed on the Test Result Monitor window).
♦
The time remaining to complete the processing of the sample.
NOTE:
If processing has completed or if there are no samples available for
processing when you select the Rack or LAS. Sample Info. button, the system
displays a message that no sample information is available.
Содержание ADVIA 2400
Страница 1: ...ADVIA 2400 Chemistry System Operator s Guide 073D0403 01 Rev A 2009 06 ...
Страница 12: ...12 ADVIA 2400 Operator s Guide ...
Страница 64: ...64 ADVIA 2400 Operator s Guide ...
Страница 98: ...98 ADVIA 2400 Operator s Guide ...
Страница 165: ...Maintenance 165 ...
Страница 167: ...Troubleshooting 167 ...
Страница 173: ...System Setup 173 ...
Страница 177: ...System Setup 177 ...
Страница 181: ...Appendix 181 ...
Страница 182: ...182 ADVIA 2400 Operator s Guide ...