Running automatic calibration
Controls requested at the Start Conditions window are run before the automatic
calibration samples.
You can use the Auto Calibration Setting dialog box at the Calibration Setup window to
request automatic calibration after a specific time interval or after a new reagent
container is loaded. If the calibration interval expires while the system is not running
samples, the automatic calibration is performed at the beginning of the next run.
However, if the calibration interval has expired and controls are requested at the Start
Conditions window, the controls are run before the automatic calibration.
At the start of a run, calibration and control samples are run using the following priority
order:
1.
Manual calibrations requested at the Start Conditions window
2.
Manual controls requested at the Start Conditions window
3.
Automatic calibrations requested on the Auto Calibration Setting dialog box
4.
Automatic controls requested on the Auto Control Setting dialog box
Using multiple reagent containers for the same test
You can use multiple, barcode-labeled reagent containers for the same test by loading the
reagent containers on Reagent Tray 1 (RTT-1) or Reagent Tray 2 (RTT-2).
After the reagent barcode labels are read, the reagent container with the lowest reagent
tray number for each test is used first. If that reagent container is empty, the reagent
container with the next lowest reagent tray number is used.
Before performing a reagent barcode read or a reagent reset, you should remove any
empty reagent containers from the reagent tray.
Содержание ADVIA 2400
Страница 1: ...ADVIA 2400 Chemistry System Operator s Guide 073D0403 01 Rev A 2009 06 ...
Страница 12: ...12 ADVIA 2400 Operator s Guide ...
Страница 64: ...64 ADVIA 2400 Operator s Guide ...
Страница 98: ...98 ADVIA 2400 Operator s Guide ...
Страница 165: ...Maintenance 165 ...
Страница 167: ...Troubleshooting 167 ...
Страница 173: ...System Setup 173 ...
Страница 177: ...System Setup 177 ...
Страница 181: ...Appendix 181 ...
Страница 182: ...182 ADVIA 2400 Operator s Guide ...