Maintenance
145
1)
Dilution Probe (DPP)
2)
Sample Probe (SPP)
3)
Reagent Probe 1 (RPP1)
4)
Reagent Probe 2 (RPP2)
Figure 5-28. Probes
Remove the probe
1.
Put the system in Standby mode.
CAUTION
To keep from damaging the probe tip when the power is off, you must manually
support the probe and be careful not to strike it against anything on the analyzer.
2.
Cover the cuvettes, wash cups, and other analyzer surfaces with lint-free towels to
catch any screws that might fall.
3.
Lift and manually rotate the probe to an accessible location.
Probe
Accessible Location
Sample probe (SPP)
Over the dilution tray (DTT)
Reagent probe 1 (RPP1)
Over reagent tray 1 (RTT1)
Reagent probe 2 (RPP2)
Over reagent tray 2 (RTT2)
4.
Loosen but do not remove the
setscrews on each side of the
probe cover. Lift the cover off
the probe.
5.
Using pliers or your fingers,
loosen the knurled fitting (
3
)
counterclockwise, then unfasten
and remove it by hand.
6.
Loosen but do not remove
setscrews (
4
)
1)
Terminal 2
2)
Joint Holder
3)
Joint Connector
4)
Philips Screws
(4 places)
5)
Probe Tube
7.
Lift the old probe and discard.
Figure 5-29. Probe without cover
Содержание ADVIA 2400
Страница 1: ...ADVIA 2400 Chemistry System Operator s Guide 073D0403 01 Rev A 2009 06 ...
Страница 12: ...12 ADVIA 2400 Operator s Guide ...
Страница 64: ...64 ADVIA 2400 Operator s Guide ...
Страница 98: ...98 ADVIA 2400 Operator s Guide ...
Страница 165: ...Maintenance 165 ...
Страница 167: ...Troubleshooting 167 ...
Страница 173: ...System Setup 173 ...
Страница 177: ...System Setup 177 ...
Страница 181: ...Appendix 181 ...
Страница 182: ...182 ADVIA 2400 Operator s Guide ...