Maintenance
135
Figure 5-22. Locating pin
CAUTION
Do not touch the glass portion of the lamp. Touching the lamp could damage it. If the
lamp is dirty, clean it using lint-free toweling moistened with ethanol.
6.
Install the lamp screws.
7.
Install the lead wires and fasten the knobs.
8.
Replace the access panel.
9.
Return the analyzer to the READY mode, if necessary.
10.
Wait 40 minutes for the lamp to stabilize.
NOTE:
You can perform the daily shutdown 2 while you are waiting.
11.
Then, check the lamp energy.
12.
Register the data and verify that attenuations register at 100%, voltages are between
5–9, and scatter is no greater than ±40.
13.
Perform the cell blank measurement test.
NOTE:
Siemens recommends the assays on the system be calibrated after the lamp is
replaced.
14.
Run QC controls to verify that all assays are within the laboratory’s established
control ranges.
Washing the ISE electrodes lines
Materials required:
• Dummy electrode with o-ring and cap
(REF 05938765, PN 073-0342-01)
• ISE detergent solution
(REF 01307361, B01-4174-01)
Time: 25 minutes
Analyzer mode: Manual operation
BIOHAZARD
Wear personal protective
equipment.
Use universal precautions.
Содержание ADVIA 2400
Страница 1: ...ADVIA 2400 Chemistry System Operator s Guide 073D0403 01 Rev A 2009 06 ...
Страница 12: ...12 ADVIA 2400 Operator s Guide ...
Страница 64: ...64 ADVIA 2400 Operator s Guide ...
Страница 98: ...98 ADVIA 2400 Operator s Guide ...
Страница 165: ...Maintenance 165 ...
Страница 167: ...Troubleshooting 167 ...
Страница 173: ...System Setup 173 ...
Страница 177: ...System Setup 177 ...
Страница 181: ...Appendix 181 ...
Страница 182: ...182 ADVIA 2400 Operator s Guide ...