System Overview
41
Test Result Monitor window
Use this window to monitor an analysis while it is running.
At the Test Result Monitor window, you can monitor the following:
•
processing status
•
The center of the window resembles the sample tray. The outer ring is the STT tray;
the inner ring is the CTT tray. Each tray position is a circle.
•
As the run continues, the circles containing samples change color. The colors indicate
the current status of each sample. The color codes are displayed at the lower left of
the window.
NOTE:
When running controls, please wait, there is a delay before the color coding
updates.
Sample test status color codes
The Test Result Monitor window uses seven color codes to represent sample status,
and the seven codes are displayed on the window. The Review/Edit window uses
these seven plus an eighth to indicate that a test for the sample was ordered but not
completed. Discrepancies in color are due to the windows’ displaying different
information. For example, a "complete" sample appears as dark blue on the Test
Result Monitor window, indicating that sampling is complete, while in the Review/
Edit window the same sample appears as pale blue, indicating that some tests are
missing.
No sample (white)
The position number does not contain a sample for this run
Untested (gray)
The run began and a workorder was created for this sample, but the sample was
not yet aspirated from the sample tray.
In process (pink)
At the Test Result Monitor window – the sample was aspirated from the sample
tray and was added to the dilution tray.
At the Review/Edit window – the sample is being analyzed or is being rerun.
Complete (blue)
At the Test Result Monitor window – the sample has been aspirated
At the Review/Edit window – analysis or the rerun for the sample is complete,
and the resultant data is produced.
Pending run (yellow)
The sample was analyzed but it must be reanalyzed. The reanalysis has not yet
started.
Содержание ADVIA 2400
Страница 1: ...ADVIA 2400 Chemistry System Operator s Guide 073D0403 01 Rev A 2009 06 ...
Страница 12: ...12 ADVIA 2400 Operator s Guide ...
Страница 64: ...64 ADVIA 2400 Operator s Guide ...
Страница 98: ...98 ADVIA 2400 Operator s Guide ...
Страница 165: ...Maintenance 165 ...
Страница 167: ...Troubleshooting 167 ...
Страница 173: ...System Setup 173 ...
Страница 177: ...System Setup 177 ...
Страница 181: ...Appendix 181 ...
Страница 182: ...182 ADVIA 2400 Operator s Guide ...