Dilution tray wash mechanism operation
At initialization, the DWUD moves to the up position. If it is already up, it is lowered,
then raised.
During each 2-second cycle on the DTT, the 10 nozzles operate as follows:
Nozzle Probe
Description
1
First nozzle
C1
Aspirates wash water.
2
Second nozzle
A1
Dispenses sample and wash water.
B1
Absorbs overflow liquid (abnormal
conditions).
3
Third nozzle
C2
Aspirates wash water.
4
Fourth nozzle
A2
Dispenses sample and wash water.
B2
Absorbs overflow liquid (abnormal
conditions).
5
Fifth nozzle
C3
Aspirates wash water.
6
Sixth nozzle
A3
Dispenses sample and wash water.
B3
Absorbs overflow liquid (abnormal
conditions).
7
Seventh nozzle
C4
Aspirates wash water.
8
Eighth nozzle
A4
Dispenses sample and wash water.
A5
Absorbs overflow liquid (abnormal
conditions).
9
Ninth nozzle
C5
Aspirates wash water.
C6
Aspirates wash water.
10
Tenth nozzle
D
Vacuums remaining liquid from cuvette.
Figure 1-25. Nozzle separation
If an assay is not running, you can operate the unit manually at the Manual Operation
window.
Содержание ADVIA 2400
Страница 1: ...ADVIA 2400 Chemistry System Operator s Guide 073D0403 01 Rev A 2009 06 ...
Страница 12: ...12 ADVIA 2400 Operator s Guide ...
Страница 64: ...64 ADVIA 2400 Operator s Guide ...
Страница 98: ...98 ADVIA 2400 Operator s Guide ...
Страница 165: ...Maintenance 165 ...
Страница 167: ...Troubleshooting 167 ...
Страница 173: ...System Setup 173 ...
Страница 177: ...System Setup 177 ...
Страница 181: ...Appendix 181 ...
Страница 182: ...182 ADVIA 2400 Operator s Guide ...