116
ADVIA 2400 Operator’s Guide
Location Position
Wash
Solution
1
CTT-15
ISE Detergent Solution
1
CTT-16 Deionized
water
1
CTT-49
10% Cuvette Wash Solution (Daily)
5% Reagent Probe Wash 3 (Weekly)
1
CTT-50 Deionized
water
1
CTT-51 Deionized
water
2
RTT1-47
Reagent Probe Wash 1
2
RTT1-48
Reagent Probe Wash 2
2
RTT1-49
10% Cuvette Wash Solution (Daily)
5% Reagent Probe Wash 3 (Weekly)
2
RTT1-50 Deionized
water
3
RTT2-47
Reagent Probe Wash 1
3
RTT2-48
Reagent Probe Wash 2
3
RTT2-49
10% Cuvette Wash Solution (Daily)
5% Reagent Probe Wash 3 (Weekly)
3
RTT2-50 Deionized
water
Figure 5-14. Location of wash solutions on the CTT and RTT trays
1.
At the Operation Panel, select
Wash
.
2.
Ensure the 10-mL tube at CTT (
1
) position #51 contains DI water.
TIP
By choosing CTT position #50 for WASH 2 and CTT position #51 for WASH 3, you
only need to refill the CTT container once.