28
ADVIA 2400 Operator’s Guide
Reagent trays
Reagent trays 1 and 2 (RTT1 and RTT2) contain reagents used for assays (positions 1-46)
and the detergents used for daily washing and contamination prevention (positions 47-
50). The reagent probes (RPP1 and RPP2) aspirate the required reagent and dispense it
into the reaction tray (RRV) cuvettes for analysis.
1)
Reagent tray 1 (RTT1)
2)
Reagent tray 2 (RTT2)
Figure 1-20. Reagent trays
Each tray has 50 positions. RTT1 contains the first reagent (R1); RTT2 contains the
second reagent (R2).
You can use any reagent container for more than one test item; a test item may require
more than one reagent container.
Each reagent tray has a barcode reader (RBC-1 and RBC-2).