Maintenance
151
CAUTION
Make a note of the bottle position on the shelf, to avoid mixing up the fluid bottles.
3.
Remove the RRV bath oil bottle (
1
).
CAUTION
Ensure that level-sensor connector (
2
) does not get wet, to avoid damaging it.
4.
Refill the bottle with RRV (reaction) bath oil.
5.
Replace the bottle on the shelf in the cabinet.
6.
Connect the RRV bath oil bottle level sensor connector (
2
) by pushing the connector
in and turning it clockwise.
7.
Insert the filter and tube, then fasten the cap.
NOTE:
Make sure that the filter holder is located at the bottom of the bottle.
Preventive cleaning of the wash station lines
Materials required:
•
5 reagent containers, empty, 70-mL
(optional)
•
Probe Wash 1
•
Probe Wash 2
•
10% Cuvette Wash solution
•
5% Probe Wash 3 solution
•
70-mL reagent container
(REF 06397121, PN 073-0373-02)
•
Wash solution labels (REF 00153468,
PN 073-0406-02)
•
Deionized water
Time: 45 minutes
Analyzer mode: READY
BIOHAZARD
Wear personal protective
equipment.
Use universal precautions.
If you experience a problem with clogs in the wash station aspiration nozzles and lines,
use this procedure to clean the WUD and DWUD wash station aspiration nozzles and
lines.
Содержание ADVIA 2400
Страница 1: ...ADVIA 2400 Chemistry System Operator s Guide 073D0403 01 Rev A 2009 06 ...
Страница 12: ...12 ADVIA 2400 Operator s Guide ...
Страница 64: ...64 ADVIA 2400 Operator s Guide ...
Страница 98: ...98 ADVIA 2400 Operator s Guide ...
Страница 165: ...Maintenance 165 ...
Страница 167: ...Troubleshooting 167 ...
Страница 173: ...System Setup 173 ...
Страница 177: ...System Setup 177 ...
Страница 181: ...Appendix 181 ...
Страница 182: ...182 ADVIA 2400 Operator s Guide ...