3. Click
Explore > Show > Show Base Peak Chromatogram.
The selections are shown in the
Start Time
and
End Time
fields.
Figure 9-6 Base Peak Chromatogram Options
4. In the
Mass Tolerance
field, type the value to indicate the mass range used to find a peak. The software
finds the peak using a value twice the typed range (± the mass value).
5. Type the intensity below which peaks are ignored by the algorithm in the
Minimum Intensity
field.
6. Type the mass that determines the beginning of the scan range in the
Minimum Mass
field.
7. Type the mass that determines the end of the scan range in the
Maximum Mass
field.
8. To set the start and end times, select the
Use Limited Range
check box and do the following:
• In the
Start Time
field, type the time that determines the start of the experiment.
• In the
End Time
field, type the time that determines the end of the experiment.
9. Click
OK
.
The BPC is generated in a new pane.
TripleTOF
®
5600/5600+ System
System User Guide
85 / 131
RUO-IDV-05-7040-B
Operating Instructions — Analyze and Explore Data