Scan Type
Use
Parameter
Type
Parameter
Location
MS/MS only,
Enhanced
The amount of time in milliseconds before
the ion pulse. The default (11 msec) is
calculated based on the TOF masses and
can be adjusted by the operator. The range
is typically 6 msec to 333 msec.
This parameter is optimized using
Instrument Optimization
wizard if the
Enhanced Ion
option is selected in the
Advanced
options. In general, the default
values do not have to be changed.
Compound
Ion Release
Delay (IRD)
3
MS/MS only,
Enhanced
This is the width, or duration, of the ion
pulse in milliseconds and is calculated based
on the IRD. The range is typically 5 to 328
msec with a default value of 10 msec.
This parameter is optimized using the
Instrument Optimization
wizard if the
Enhanced Ion
option is selected in the
Advanced
options. In general, the default
values do not have to be changed.
Compound
Ion Release
Width (IRW)
3
All
The CEM parameter controls the voltage
applied to the detector. The voltage affects
the detector response.
Detector
MCP (CEM)
4
TripleTOF
®
5600/5600+ System
System User Guide
59 / 131
RUO-IDV-05-7040-B
Operating Instructions — Acquisition Methods