CAUTION: Potential Mass Shift. Maintain a stable ambient temperature. If the temperature
changes by more than 2 °C per hour, then the resolution and mass calibration might be
affected.
Electromagnetic Environment
Electromagnetic Compatibility
Basic Electromagnetic Environment:
Environment existing at locations characterized by being supplied
directly at low voltage from the public mains network.
Performance Criteria A (Criteria A):
Equipment shall operate as intended with no degradation of performance
and no loss of function during or after test.
Performance Criteria B (Criteria B):
Equipment may experience loss of function (one or more) during test
but shall operate as intended with some degradation of performance and functions self-recoverable after test.
Performance Criteria C (Criteria C):
Equipment may experience loss of function (one or more) during test
but shall operate as intended with some degradation of performance and functions recoverable by operator after
test.
The equipment is intended for use in a basic electromagnetic environment.
The expected performance loss under the electromagnetic immunity conditions is less than 20% change in total
ion count (TIC).
Make sure that a compatible electromagnetic environment for the equipment can be maintained so that the device
will perform as intended. If the power supply line is subject to high electrical noise, then install a surge protector.
Electromagnetic Interference
Class A Equipment:
Equipment which is suitable for use in all establishments other than domestic and those
directly connected to a low voltage power supply network which supplies buildings used for domestic purposes.
[Derived from CISPR 11:2009, 5.3] Class A equipment shall meet Class A limits.
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part
15 of the FCC (Federal Communications Commission) Compliance Rules.
These limits are designed to provide reasonable protection against harmful interference when the equipment is
operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy
and, if not installed and used in accordance with the operator's manual, can cause harmful interference to radio
communications.
Operation of this equipment in a residential area is likely to cause harmful interference in which case you will be
required to correct the interference, at your own expense. Changes or modifications not expressly approved by
the manufacturer could void your authority to operate the equipment.
TripleTOF
®
5600/5600+ System
System User Guide
13 / 131
RUO-IDV-05-7040-B
Operational Precautions and Limitations