Table 9-3 Right-click Menu for Chromatographic Peak List Tab
Function
Menu
Show the peaks in two colors in the graph.
Show Peaks in Graph
Opens the
Intelliquan
dialog.
IntelliQuan Parameters
Saves the data as a txt file.
Save As Text
Deletes the selected pane.
Delete Pane
Show ADC Data
ADC (analog-to-digital converter) data is acquired from a secondary detector (for example from a UV detector
through an ADC card), and is useful for comparison with mass spectrometer data. To have ADC data available,
acquire the data and the mass spectrometer data simultaneously and then save it in the same file.
1. Make sure that the
Example
folder is selected.
2. On the Navigation bar, under
Explore
, double-click
Open Data File
.
The Select Sample dialog opens.
3. In the
Data Files
field, double-click
Devices
and then click
Adc16chan.wiff
.
4. In the
Samples
list, select a sample, and then click
OK
.
5. Click
Explore > Show > Show ADC Data
.
Figure 9-2 Select ADC Channel Dialog
6. In the
Channel
list, select a channel, and then click
OK
.
The ADC data is shown in a new pane beneath the active pane.
System User Guide
TripleTOF
®
5600/5600+ System
RUO-IDV-05-7040-B
78 / 131
Operating Instructions — Analyze and Explore Data